Troubleshooting Immunoperoxidase Assays


Problem: Positive Control is negative or very weak.

 

Possible Causes:


1. Please note that Controls available from Fuller Laboratories have been bottled at "screening dilution". This dilution depends upon the specific assay, but in all cases Positive Controls are intended to be used initially without dilution to demonstrate a "strong" reaction. These Controls give endpoint titers 8-fold beyond the screening dilution (range 1:4-1:16).
   
2. If IFA Substrate Slides are purchased as components, the Positive and Negative Controls available from Fuller Laboratories may be used to locate the optimal conjugate titer for use. This optimal titer should give a clear 1+ positive reaction at a Positive Control dilution of 1:8 (see #1 above) and a clearly negative reaction using the undiluted Negative Control.